Triple Plus+Triple Plus+ IRUser ManualM07701February 2010Issue 4a
6Operation Triple Plus+ and Triple Plus+ IRlevel by pressing the UNMARKED button. The peak gas level will then be displayed again from then on. In ad
7Triple Plus+ and Triple Plus+ IR Operationthe whole display with the warning. All instantaneous alarm types are latch-ing which means
8Setting Up Triple Plus+ and Triple Plus+ IRIII. Setting Up The Triple Plus+/Plus+ IR has been designed to display as much or as little infor-mation
9Triple Plus+ and Triple Plus+ IR Setting UpHow to change an optionTo change an option, use the first two keys, labelled with the << and >&g
10Setting Up Triple Plus+ and Triple Plus+ IR3.2 Configuration Options LIGHT:Configures the backlight timeout option to be NEVER TIMEOUT, 10
11Triple Plus+ and Triple Plus+ IR Setting UpZERO enables the option of zeroing the sensors at start, as per the zero all option above
12Setting Up Triple Plus+ and Triple Plus+ IRmodules. The exact list of gases and their corrections are configurable via a PC using Portables PC soft
13Triple Plus+ and Triple Plus+ IR Setting UpIMPORTANT SAFETY NOTICE: All units fitted with sensors to measure both %LEL AND %Volume Methane, other h
14Setting Up Triple Plus+ and Triple Plus+ IROFF Two options: OFF ENABLED, OFF DISABLED. Determines whether the instru-ment can be switched
15Triple Plus+ and Triple Plus+ IR Quickcal3.3 QuickcalThe instrument has the ability to perform a ‘quick calibration’ of four gas chan-nels: CH4; CO
Instructions specific for use in hazardous areasThe following instructions apply to equipment cov-ered by certificate numbers: Sira 02ATEX2176X IECEx
16Pellistor saving Triple Plus+ and Triple Plus+ IRAdditionally, a pellistor will switch itself off if it reaches a reading of 105% LEL or more – thi
17Triple Plus+ and Triple Plus+ IR Configuration optionstor, and the actual correction factor itself may be modified (Pellistor only). INSTANTANEOUS
18Battery Charging Triple Plus+ and Triple Plus+ IRIV. Battery ChargingThe new LIBRA Lithium-ion battery pack must be charged for a full 12 hours bef
19Triple Plus+ and Triple Plus+ IR Battery ChargingIf the instrument is to remain unused for a length of time, it should be charged prior to the peri
20Data Logging Triple Plus+ and Triple Plus+ IRV. Data LoggingThe data logging aspect of the instrument can be configured to be active when the inst
21Triple Plus+ and Triple Plus+ IR Troubleshooting GuideVI. Troubleshooting GuideSYMPTOM DIAGNOSIS REMEDY/CHECKDoes not switch on Battery flat Re
22Maintenance and Calibration Triple Plus+ and Triple Plus+ IRVII. Maintenance and CalibrationCrowcon recommends monthly gas response checks and a re
23Triple Plus+ and Triple Plus+ IR Maintenance and Calibration7.3 Fitting the Lithium-ion Battery PackThe Li-ion battery pack may be used with ATEX/I
24Limitations of Use Triple Plus+ and Triple Plus+ IRVIII. Limitations of UseLONG TERM STORAGE LIMITS OPERATING LIMITS min max min maxTEMPERATURE 0°C
25Triple Plus+ and Triple Plus+ IR Accessories and sparesIX. Accessories and sparesAccessories list Crowcon Description part numberCarrying case a
LRQAISO9001:2008.II 2G: Ex ib d IIC T4 Gb
26Accessories and spares Triple Plus+ and Triple Plus+ IR0– 100 ppm nitric oxide S012450– 25% vol oxygen S0113710– 10ppm hydrogen chloride* S017
27Triple Plus+ and Triple Plus+ IR Appendix 1 – Pumped version (optional)Appendix 1 – Pumped version (optional)The internal pump, when used in conjun
28Appendix 2 – Default Flammable Correction Factors Triple Plus+ and Triple Plus+ IRAppendix 2 – Default Flammable Correction FactorsThis information
29Triple Plus+ and Triple Plus+ IR Appendix 3 – Front Panel Text ItemsAppendix 3 – Front Panel Text ItemsThis Appendix lists all the possible values
30Appendix 3 – Sensor Limitations Triple Plus+ and Triple Plus+ IR**Only if MUTE function enabled (via Portables PC). LOG level configuration items:
31Triple Plus+ and Triple Plus+ IR Appendix 4 – Sensor LimitationsAppendix 4 – Sensor LimitationsThe instrument is not suitable for use in ambient te
32Appendix 5 – Menu Map Triple Plus+ and Triple Plus+ IRAppendix 5 – Menu Map15 APPENDIX 5 – Menu Map2424This function does not appear in Triple Plus
ContentsQuickstart Guide ... 1I. General Description ...
UK OfficeCrowcon Detection Instruments Ltd2 Blacklands Way, Abingdon Business ParkAbingdon Oxfordshire OX14 1DY United KingdomTel: +44 (0)1235 557700
1Triple Plus+ and Triple Plus+ IR Quickstart Guide1Quickstart GuidePrior to first use, the LIBRA Lithium-ion battery pack requires a full 12-hour cha
2General description Triple Plus+ and Triple Plus+ IRI. General DescriptionTriple Plus+The Crowcon Triple Plus+ is a microprocessor controlled portab
3Triple Plus+ and Triple Plus+ IR General description(b) No damage caused by exposure to high gas concentrations (c) No poisoning effects (d) Abilit
4Operation Triple Plus+ and Triple Plus+ IRII. Operation2.1 Switch OnPress the 'ON' button to switch on the Triple Plus+/Plus+ IR
5Triple Plus+ and Triple Plus+ IR Operation2.2 DisplayThe information on the display is dependent on the type of sensors installed in the instrument.